Part Number Hot Search : 
TIP31A HZZ00113 FMG23S MBZ5230 85NS160 DSP16318 MBR1045 HER103
Product Description
Full Text Search
 

To Download 3239 Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 DIGITAL HiTESTER 3239 (4-terminal function) 3238 (Advanced model) 3237 (Economically priced)
Field measuring instruments
Outstanding performance for production lines with a sampling rate of 3.3 ms
High-speed DMM
The DIGITAL HiTESTERs 3237, 3238, and 3239 can perform 3.3 ms high-speed sampling, and come equipped with a comparator, external input and output, and an RS-232C interface. These three highperformance DMMs can be used not only in laboratories, but in production lines that require the minimal tact time. The 3237 is the basic model, and is equipped with the basic necessary functions. The 3238 is a highprecision, broadband model that also features current measurement terminals and a frequency measurement function. The 3239 includes the functions of the 3238 plus the 4-terminal resistance measurement function. All three units are designed with emphasis on measurement speed and safety.
1
3.3 ms/sample High-speed Performance and Reliability
Features
Sampling speed Values in the ( ) show samples/second. Frequency 50 Hz 60 Hz FAST* MEDIUM SLOW 3.3 1 ms (300) 130 5 ms (8) 1,040 50 ms (1) 3.3 1 ms (300) 108 5 ms (9) 1,080 50 ms (1)
Samples at rates of up to 300 samples/sec. (3.3 ms/sample) Comparator function provides high-speed pass/fail evaluation Equipped with external input and output for sequence control Usefull Save/Load function helps work go faster
The 3237, 3238 and 3239 are equipped with a variety of functions that help minimize tact time in production lines.
For details, see page 2.
Does not apply at resistances higher than 2M, or LP higher than 200k (see page 5). For the 3238 and 3239's frequency function gate time, see page 5.
* Approximately 55 ms required for self-calibration at 30-minute intervals.
True RMS value measurement
Both the 3237 and 3238 use true RMS measurement for determination of distorted waveforms. In fact, HIOKI guarantees accuracy of the 3238 and 3239 for AC voltage of 10 Hz to 300 kHz, and AC current of 10 Hz to 30 kHz.
For specifications, see pages 5 and 6.
Low power resistance measurement function prevents sample deterioration
The 3237, 3238 and 3239 use a low power function to minimize sample degradation when measuring resistance. With this function, open terminal voltage never goes over 0.45 V DC, and measurement current never surpasses 100 A DC.
For specifications, see pages 5 and 6.
Interface supports full remote operation
Measurement can be automated by using a controller to operate the 3237 or 3238 through the GP-IB or RS-232C interface.
For details, see page 3.
Select from 3 types of units
The basic and economical
3237
32 37 32 38 32 39
The high-accuracy & multi-functional
3238 3239
DC V basic accuracy: 0.025% rdg.2dgt.
DC V basic accuracy: 0.01% rdg.2dgt. Includes frequency measurement for AC and DC A
DC voltage AC voltage
[5 ranges, 199.999 mV to 1000.00 V] [4 ranges, 1999.99 mV to 700.00 V] 2-terminal [7 ranges, 199.999 to 100.000 M] 2-terminal [4 ranges, 1999.99 to 1999.99 M]



For 4-terminal resistance measurement
Resistance
DC V basic accuracy: 0.01% rdg.2dgt. All the functions of the 3238, plus 4-terminal measurement Reliable resistance measurement using the 4-terminal measurement method
Using 4-terminal resistance measurement, which is unaffected by variables such as measurement lead wiring resistance, the 3239 displays outstanding resistance measurement capabilities.
Resistance LP
Continuity check Diode check
[A buzzer sounds when resistance is less than 50.00 ] [Anode-cathode voltage in the 1999.99 mV range]
Current measurement by clamp sensor AC/DC current Frequency
4-terminal 4-terminal
[2 ranges, 199.999 mA and 1999.99 mA] [5 ranges, 99.9999 Hz to 300.000 kHz]
Resistance
[5 ranges,199.999 to 1999.99 k]
Resistance LP
[4 ranges,1999.99 to 1999.99 M] For clamp specifications, see page 4 For DIGITAL HiTESTER specifications, see pages 5 and 6
2
Minimizing tact time with sequence control
High-speed comparator and external input/output
Connector used: 57RE-40360-730B (D29) (DDK Ltd.)
EXT.I/O Loading of saved settings from panel......LOAD 0 to 4 Measurement start trigger input............TRIG Measurement end signal output............EOC Comparator output..............................Hi, IN, Lo Internal power supply +5 V (max. 50 mA)......INT. DC V Internal GND........................................INT. GND Comparator with external output
You can set the upper and lower limits, and display one of 3 results: Hi, IN or Lo. In addition to LED and buzzer results, open collector output results are provided through the external input/output terminals. X: measurement value, H: Upper limit, L: Lower limit X > H..............................Hi H > X > L.....................IN L > X..................Lo
Conforming connector: ADS-HC360001-010 (Honda Tsushin Kogyo Co., Ltd.), RC30-36P (Hirose Electric Co., Ltd.) and other suitable connectors Input type: CMOS level Signal voltage: Hi 3.8 V to 5.0 V, Lo 0 V (shorted) to 1.2 V Logic: Negative logic (Low active) Output type: Open collector Applied voltage: Max. 35 V DC, 50 mA Logic: Negative logic (Low active)
With free running measurement
MIN.
Time TYP. ... See below
MAX. ...
With external control
T1 Measurement trigger pulse width 500 s T2 T3 T4
Trigger delay time Sampling time using external control
FAST See the table MEDIUM at the top right of page 1. SLOW ... 2.0 ms ... ... ...
Internal operation time
the of measurement 500 s T5 Fromuntil end next trigger the
... 1.7 ms ... EOC Lo level time FAST 50 ms ... T6 for free running MEDIUM ... measurement SLOW ... 500 ms ... FAST Sampling time See the table T7 for free running MEDIUM at the top right of page 1. measurement SLOW
Save/Load function for rapid response to various work situations
You can save and recall a maximum of 30 DMM setting conditions for various range and comparator values.
A trigger delay designed for measurement safety
The 3237, 3238 and 3239 are equipped with a trigger delay function that can be set to manual or automatic for the time period between trigger input and the display of the comparator result (see T2 in the figure above).
Manual settings: Designate periods in terms of millisecond intervals between 0.000 s and 9.999 s Automatic settings: FAST MEDIUM SLOW
DC V AC V
(200 to 200 k)
3 ms 500 ms 3 ms
3 ms 800 ms 3 ms
3 ms 1.5 s 3 ms
3
Automation of Line Inspection
Available interfaces
GP-IB (option -01 specifications)
Purpose: Remote control and measurement value output
Standards conformance : IEEE -488.1 1987 Reference standard Transmission speed
(reference data)
Power line frequency: 60Hz TRIG: EXT.Trig Command: [:READ ?]
RS-232C (standard)
Purpose: Remote control and measurement value output
Transmission system Transmission speed Data bit length Stop bits Parity bits Delimiter Handshaking XON/XOFF Connector : Asynchronous method Full duplex : 9600 bps (fixed) : 8 bits :1 : None : CR+LF : Hardware : Not used : 9-pin D-sub connector
: IEEE -488.2 1987
FAST MEDIUM SLOW Transmission 1,080 7.0 ms 108 ms ms speed
Controller: PC-9801 RA (NEC)
OS: MS-DOS Ver. 3.30, N88-BASIC Ver. 6.0
Interface function: SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, CO User code User connector
: ASCII code : 24-pin IEEE488 interface bus connector
All functions except switching the power on and off can be completely remote controlled and measurement data collected via either the GP-IB or RS-232C interface
Please inquire regarding compatibility with the command sets of other manufacturers.
Output data to a printer (option)
When an RS-232C compatible PRINTER 9442 is connected, you can print measurements by pressing the key if in manual trigger mode, or the key if in internal trigger (free run) mode.
Item No. Judgment
PRINTER 9442
CONNECTOR CABLE 9444
Cord length approx 1.5m
AC ADAPTER 9443
Measurement function
Measurement value 9443-02 9443-01 (For the EU) (For Japan) 9443-03 (For the America)
Equipped with a foot switch for printer control
As an alternative to pressing the key or the key, you can also connect a foot switch to the external I/O TRIG terminal. You can then initiate printing by stepping on the foot switch (closing the circuit).
: Thermal serial dot matrix number suffix when ordering. : 112 mm : 52.5cps : AC ADAPTER 9443 or supplied nickel-hydride battery (capable of printing about 3000 lines on full charge from 9443) Dimensions and mass : Approx. 160W 66.5H 170D mm; approx. 580 g Printing method Paper width Printing speed Power supply
Please specify appropriate model
When you purchase a PRINTER 9442, you must also purchase a CONNECTION CABLE 9444 and a AC ADAPTER 9443 to connect it to the DMM.
4
Efficient Evaluation Testing
PC measurement using the high accuracy and broad coverage of the 3238 and 3239
Time Current Integration
Example of data processing with Microsoft Excel
Time History Graph of Current Integration
Time
Battery Discharge Characteristics
Highly accurate measurement with minimal drift
The unit uses self-regulation to suppress drift. Also, the DMM is ideal for collecting data over extended periods of time.
Using Excel for efficient data processing
The DMM supports fast data processing by allowing you to transfer data directly to a worksheet through either the GP-IB or RS-232C interface.
Consult your nearest HIOKI dealer for details on software
Supports large AC current measurement by clamp sensor
Easy setup Both the 3237, 3238 and 3239 can measure live line currents using an optional clamp sensor. Enter the name of the clamp sensor being used and display current values simply by selecting a range.
Clamp sensor settings screen
Select 9010
Select the 20 A range.
From the menu's clamp sensor selection screen, select the name of the sensor with the cursor key and press the key. Then, select the same range as you set for the sensor with the cursor key.
* The accuracy of the clamp sensors shown on the left (when used with the DMM) is calculated by taking: the difference in the AC V accuracy for the DMM (dgt.) 10 (dgt.). For the AC V accuracy of the DMM, see page 6. In addition to the sensors described above, you can also connect the 3283, 3284, 3285 (requires the 9094) Clamp On HiTesters, and the 9277, 9278, 9279 (requires the 9555) DC Sensors.
5
3237, 3238, 3239 common specifications (Accuracy at 23C5C (73F9F), 80% rh or less)
DC voltage (DC V) Range Resolution Full scale Input impedance Overload protection 200 mV 1 V 199.999mV Greater than 100M 1000 V DC 2000 mV 10 V 1999.99mV Greater than 100M 750 V AC 20 V 100 V 19.9999 V Appox. 11 M However, less than 200 V 1 mV 199.999 V Appox. 10 M 107 V Hz 1000 V 10 mV 1000.00 V Appox. 10 M Resistance () 2-terminal measurement Range Resolution Full scale Current Open terminal voltage Overload protection 200 1 m 199.999 Appox. 1 mA 6V DC max. 2000 10 m 1999.99 Appox. 1 mA 6V DC max. 20 k 100 m 19.9999k Appox. 100A 6V DC max. 200 k 1 199.999k Appox. 10A 6V DC max. 500Vpeak 2000 k 10 1999.99k Appox. 1 A 6V DC max. 20 M 100 19.9999M Appox. 100nA 6V DC max. 100 M 1 k 100.000M Appox. 20nA 6V DC max.
For fast sampling in the 20 M range or higher. Frequency FAST*
For sampling at in the 2 M range or the LP 200 k.range or higher
AC voltage (AC V) Range Resolution Full scale 2000 mV 10 V 1999.99mV 20 V 100 V 19.9999 V 200 V 1 mV 199.999 V 700 V 10 mV 750.00 V Input impedance Overload protection Appox. 1 M 600 V DC Appox. 1 M 750 V rms, 1000Vpeak Appox. 1 M However, less than 107 V Hz Appox. 1 M
Resistance () at Low Power function 2-terminal measurement Range Resolution Full scale 2000 10 m 1999.99 20 k 100 m 19.9999k 200 k 1 199.999k 2000 k 10 1999.99k Continuity check Range Resolution Full scale Current Open terminal voltage Overload protection 2000 10 m 1999.99 Appox. 100A 0.45V DC max. 500 Vpeak
A built-in buzzer sounds when the resistance value is less than 50.00 .
Current Open terminal voltage Overload protection Appox. 100A 0.45V DC max. Appox. 10A 0.45V DC max. 500Vpeak Appox. 1 A 0.45V DC max. Appox. 100nA 0.45V DC max.
MEDIUM
SLOW
Frequency
FAST*
50 Hz 20 1 ms 170 5 ms 1,360 50 ms 60 Hz 16.7 1 ms 142 5 ms 1,420 50 ms
* Approximately 55 ms required for self-calibration at 30-minute intervals.
50 Hz 20 1 ms 60 Hz 16.7 1 ms
Diode check Range Resolution Full scale Current Open terminal voltage Overload protection 2000 mV 10 V 1999.99mV Appox. 1 mA 6V DC max. 500 Vpeak
3238, 3239 specifications (Accuracy at 23C5C (73F9F), 80% rh or less)
AC/DC current (A) Range Resolution Full scale Internal resistance Overload protection 200 mA 1 A 199.999mA Appox. 1 250V, 2A fuse 2000 mA 10 A 1999.99mA Appox. 100 m AC current (AC A) 200mA range Accuracy %, ppm=reading error, d=digit error Range Frequency DC current (DC A) Accuracy %, ppm=reading error, d=digit error Range 200 mA 2000 mA Thermal coefficient 0.1 %20d 400ppm20d 300ppm10d 300ppm10d 300ppm10d 500ppm30d 0.1 %30d
Sampling
SLOW 0.1 %6d 0.15%6d MEDIUM 0.1 %10d 0.15%10d FAST 0.1 %300d 0.15%300d
Thermal coefficient 100ppm0.6d 150ppm0.6d Thermal coefficient 0.12%20d 600ppm20d 400ppm10d 400ppm10d 600ppm20d 0.12%30d undefined
AC current (AC A) 2000mA range Accuracy
Sampling
SLOW 1.0%200d 0.4%200d 0.3%100d 0.3%100d 0.3%100d 0.5%300d 1.0%300d MEDIUM undefined undefined 0.5%200d 0.4%200d 0.4%200d 0.5%300d 1.0%300d FAST undefined undefined undefined 0.4%300d 0.4%300d 0.5%400d 1.0%400d
Sampling
SLOW 1.2%200d 0.6%200d 0.4%100d 0.4%100d 0.6%200d 1.2%300d undefined MEDIUM undefined undefined 0.6%200d 0.6%200d 0.6%200d 1.2%300d undefined FAST undefined undefined undefined 0.6%300d 0.6%300d 1.2%400d undefined
10 Hz to 20 Hz 20 Hz to 45 Hz 45 Hz to 300 Hz All Ranges 300 Hz to 1 kHz 1 kHz to 3 kHz 3 kHz to 10 kHz 10 kHz to 30 kHz
Specified input is 16 mA or higher Specified input is 160 mA or higher Additional error due to crest factor: 1 Frequency (Hz) Source is AC V only and input level is higher than 8% of full scale Range Resolution Full scale Internal resistance Min. measurement Overload protection 100 Hz 0.1 mHz 99.9999 Hz Appox. 1M 10 Hz 600 V DC 1 kHz 1 mHz 999.999 Hz Appox. 1M 10 Hz 750 V rms, 10 kHz 10 mHz 9.99999kHz Appox. 1M 10 Hz 1000Vpeak 100 kHz 100mHz 99.9999kHz Appox. 1M 10 Hz However, less than 107 V Hz 300 kHz 1 Hz 999.999kHz Appox. 1M 10 Hz
Frequency (Hz) Accuracy %, ppm=reading error, d=digit error For all gate times Thermal Square-wave input between 10 Hz to 300 kHz, 10 V p-p. coefficient 0.015% 2d 5 ppm All Ranges Range Frequency gate time
FAST MEDIUM SLOW
15 6 ms
110 10 ms
1,010 20 ms
Measurement time: from gate time to the input signal period 2
3239 specifications (Accuracy at 23C5C (73F9F), 80% rh or less)
Resistance () 4-terminal measurement Range Resolution Full scale Current Open terminal voltage Overload protection 200 1 m 199.999 Appox. 1 mA 6V DC max. V, terminal 2000 10 m 1999.99 Appox. 1 mA 6V DC max. 500Vpeak 20 k 100 m 19.9999k Appox. 100A 6V DC max. 200 k 1 199.999k Appox. 10A 6V DC max. SENSE terminal 400Vpeak 2000 k 10 1999.99k Appox. 1 A 6V DC max. Resistance () at Low Power function 4-terminal measurement Range Resolution Full scale 2000 10 m 1999.99 20 k 100 m 19.9999k 200 k 1 199.999k 2000 k 10 1999.99k Current Open terminal voltage Overload protection Appox. 100A 0.45V DC max. V, terminal Appox. 10A 0.45V DC max. 500Vpeak Appox. 1 A 0.45V DC max. SENSE terminal Appox. 100nA 0.45V DC max. 400Vpeak
6
3237 DC voltage (DC V) Accuracy %, ppm=reading error, d=digit error Range 200 mV 2000mV 20 V 200 V 1000 V
3238, 3239 DC voltage (DC V) Accuracy %, ppm=reading error, d=digit error
Sampling
SLOW 0.026%6d 0.025%2d 0.028%5d 0.028%2d 0.028%2d MEDIUM 0.026%10d 0.025%8d 0.028%10d 0.028%8d 0.028%8d FAST 0.035%300d 0.03%100d 0.035%100d 0.035%100d 0.035%100d
Thermal coefficient 20ppm0.6d 15ppm0.2d 20ppm0.5d 20ppm0.2d 20ppm0.2d
Sampling
SLOW 0.012%6d 0.01 %2d 0.016%5d 0.016%2d 0.016%2d MEDIUM FAST 0.012%10d 0.02%300d 0.01 %8d 0.015%100d 0.016%10d 0.02%100d 0.016%8d 0.02%100d 0.016%8d 0.02%100d
Thermal coefficient 12ppm0.6d 10ppm0.2d 16ppm0.5d 16ppm0.2d 16ppm0.2d
CMRR (50/60Hz Rl=1k): SLOW 130dB, MEDIUM 90dB, FAST 20dB
NMRR (50/60Hz ): SLOW 70dB, MEDIUM 50dB, FAST 0dB
3237 AC voltage (AC V) Accuracy %, ppm=reading error, d=digit error Range Frequency
3238, 3239 AC V Accuracy %, ppm=reading error, d=digit error
Sampling Thermal coefficient SLOW MEDIUM FAST SLOW 10 Hz to 20 Hz 1.5%200d undefined undefined 0.15%20d 0.8%200d 20 Hz to 45 Hz 0.5%200d undefined undefined 500ppm20d 0.2%200d 45 Hz to 300 Hz 0.2%100d 0.5%300d undefined 200ppm10d 0.1%100d 300 Hz to 3 kHz 0.2%100d 0.2%200d 0.2%300d 200ppm10d 0.1%100d All 3 kHz to 10 kHz 0.3%200d 0.3%200d 0.3%300d 300ppm20d 0.1%100d Ranges 10 kHz to 30 kHz 1.5%600d 1.5%600d 1.5%700d 0.15%60d 0.3%400d 30 kHz to 50 kHz undefined undefined undefined undefined 0.3%400d 50 kHz to 100kHz undefined undefined undefined undefined 1.5%1000d 100kHz to 300kHz undefined undefined undefined undefined 5.0%5000d The accuracy above is standard for inputs higher than 8% of full scale (higher than 160 V for a range of 750 V).
3237 Resistance () Accuracy %, ppm=reading error, d=digit error Measurement Range
Sampling
MEDIUM undefined undefined 0.3%200d 0.1%200d 0.1%200d 0.3%400d 0.3%400d 1.5%1000d 5.0%5000d FAST undefined undefined undefined 0.1%300d 0.1%300d 0.3%500d 0.3%500d 1.5%1100d 5.0%5000d
Thermal coefficient 800ppm20d 200ppm20d 100ppm10d 100ppm10d 100ppm10d 300ppm40d 300ppm40d 0.15%100d 0.5%500d
Additional error due to crest factor: 13239), 3 3238, 3239 Resistance () Accuracy %, ppm=reading error, d=digit error Thermal coefficient 50ppm0.8d 50ppm0.2d 50ppm0.2d 50ppm0.2d 50ppm0.2d 300ppm0.4d 0.3%1d
Sampling
SLOW 0.05 %8d 0.05 %2d 0.05 %2d 0.05 %2d 0.05 %2d 0.3 %4d 3.0 %10d MEDIUM 0.05 %18d 0.05 %12d 0.05 %12d 0.05 %12d 0.05 %12d 0.3 %20d 3.0 %50d FAST 0.05%300d 0.05%100d 0.05%100d 0.05%200d 0.05%200d 0.3 %200d 3.0 %500d
Sampling
SLOW 0.03 %8d 0.02 %2d 0.02 %2d 0.02 %2d 0.03 %2d 0.2 %4d 3.0 %10d MEDIUM 0.03 %18d 0.02 %12d 0.02 %12d 0.02 %12d 0.03 %12d 0.2 %20d 3.0 %50d FAST 0.03%300d 0.02%100d 0.02%100d 0.02%200d 0.03%200d 0.2 %200d 3.0 %500d
200 2000 20 k 2terminal 200 k measurement 2000 k 20 M 100 M
Thermal coefficient 30ppm0.8d 20ppm0.2d 20ppm0.2d 20ppm0.2d 30ppm0.2d 200ppm0.4d 0.3%1d
After zero adjustment. When measuring high resistance, use a shielded cable such as the 9236 CONNECTION CORD (1.7m).
3237 Resistance () Accuracy at Low Power function Measurement Range
3238, 3239 Resistance () Accuracy at Low Power function Thermal coefficient 50ppm0.6d 50ppm0.6d 50ppm0.6d 300ppm0.6d
Sampling
SLOW 0.05 %6d 0.05 %6d 0.05 %6d 0.3 %6d MEDIUM 0.05 %14d 0.05 %14d 0.05 %14d 0.3 %20d FAST 0.05 %300d 0.05 %300d 0.05 %300d 0.3 %500d
Sampling
SLOW 0.02 %6d 0.02 %6d 0.02 %6d 0.2 %6d MEDIUM 0.02 %14d 0.02 %14d 0.02 %14d 0.2 %20d FAST 0.02%300d 0.02%300d 0.02%300d 0.2 %300d
2000 20 k terminal 200 k measurement 2000 k
2-
Thermal coefficient 20ppm0.6d 20ppm0.6d 20ppm0.6d 200ppm0.6d
After zero adjustment. When measuring high resistance, use a shielded cable such as the 9236 CONNECTION CORD (1.7m).
3237 Continuity check Accuracy %, ppm=reading error, d=digit error Range 2000
3238, 3239 Continuity check Accuracy %, ppm=reading error, d=digit error
Sampling
FAST only 0.05 %300d
Thermal coefficient 50ppm0.6d
Sampling
FAST only 0.02 %300d
Thermal coefficient 20ppm0.6d
3237 Diode check Accuracy %, ppm=reading error, d=digit error Range 2000
3238, 3239 Diode check Accuracy %, ppm=reading error, d=digit error
Sampling
SLOW 0.025% 2d MEDIUM 0.025% 8d FAST 0.03% 100d
Thermal coefficient 15ppm0.2d
Sampling
SLOW 0.01 %2d MEDIUM 0.01 %8d FAST 0.015%100d
Thermal coefficient 10ppm0.2d
Resistance () Accuracy %, ppm=reading error, d=digit error Measurement Range
4-terminal measurement
Resistance () Accuracy at Low Power function Thermal coefficient 30ppm0.8d 20ppm0.2d 20ppm0.2d 20ppm0.2d 30ppm0.2d
4-terminal measurement
Sampling
SLOW 0.03 %8d 0.02 %2d 0.02 %2d 0.02 %2d 0.03 %2d MEDIUM 0.03 %18d 0.02 %12d 0.02 %12d 0.02 %12d 0.03 %12d FAST 0.03 %300d 0.02 %100d 0.02 %100d 0.02 %200d 0.03 %200d
Sampling
SLOW No range 0.02 %6d 0.02 %6d 0.02 %6d 0.2 %6d MEDIUM No range 0.02 %14d 0.02 %14d 0.02 %14d 0.2 %20d FAST No range 0.02%300d 0.02%300d 0.02%300d 0.2 %300d
200 2000 terminal 20 k measurement 200 k 2000 k
4-
Thermal coefficient No range 20ppm0.6d 20ppm0.6d 20ppm0.6d 200ppm0.6d
The accuracy quoted above is for a contact resistance of 100 or less.
7
3237, 3238, 3239 General Specifications
AC measurement: True RMS value measurement Crest factor: 3.0 max. Ancillary functions: Comparator, Average (0 to 99 times), Zero Adjust, Trigger (the display changes when the trigger is activated), and the Save/Load functions. (Up to 30 types of setting conditions) Interface: External input/output, RS-232C and GP-IB (option -01 specifications) Display: LED max. 199999 (999999 for frequency) Sampling rate (see page 1): SLOW approx. 1 samples/s MEDIUM approx. 8 to 9 samples/s FAST approx. 300 samples/s (Does not apply at resistances higher than 2M, or LP higher than 200k)
(self-calibration takes place for approximately 55 ms at 30-minute intervals for FAST sampling only.)
Range selection: Auto and Manual Applicable standards: Safety: EN61010-1+A2, EN61010-2-031 Lo terminal: CAT I (500V), CAT II (300V) Hi terminal: CAT I (1000V), CAT II (600V) EMC: EN61326+A1 Class B, EN61000-3-2+A1+A2, EN61000-3-3 Ambient temperature of use: 0 to 40 C(32F to 104F) 80%RH (no condensation) Storage temperature range: -10 to 50C(-14F to 122F) 70%RH (no condensation) Power supply: Select from AC 100 V/120 V/220 V/240 V, (50/60 Hz) specify when ordering Maximum rated power: 15 VA Dimensions and mass: Approx. 215 W80 H265 D mm, 2.6kg
Approx 8.5 W 3.5 H 10.4 D, 91.7 oz.
Economical Type
Options
RS-232C cable RS-232C CABLE 9637 (9pin-9pin, Reverse type/1.8m) RS-232C CABLE 9638 (9pin-25pin, Reverse type/1.8m) GP-IB cable The specifications of the 3237, 3238 and 3239 are -01 specifications GP-IB CABLE 9151-02 (2m) GP-IB CABLE 9151-04 (4m) Printer PRINTER CONNECTOR CABLE AC ADAPTER AC ADAPTER AC ADAPTER RECORDING PAPER 9442 9444 (For 9442 printer) 9443-01 (For 9442 printer, Japan) 9443-02 (For 9442 printer, EU) 9443-03 (For 9442 printer, America) 1196 (For printer, 10 rolls)
DIGITAL HiTESTER 3237 DIGITAL HiTESTER 3237-01 (with GP-IB)
Advanced Type
DIGITAL HiTESTER 3238 DIGITAL HiTESTER 3238-01 (with GP-IB)
4-terminal function & Advanced Type
DIGITAL HiTESTER 3239 DIGITAL HiTESTER 3239-01 (with GP-IB)
When you purchase a PRINTER 9442, you must also purchase a 9444 CONNECTOR CABLE and a AC ADAPTER 9443 to connect it to the DMM. cord length 65cm
TEST LEAD 9170-10 (included) Options
Clamp sensors CLAMP ON PROBE 9010-50 (10/20/50/100/200/500A AC ) CLAMP ON PROBE 9018-50 (10/20/50/100/200/500A AC, Broadband type) CLAMP ON PROBE 9132 (20/50/100/200/500/1000A AC )
For Clamp sensors specifications, see page 4.
9010-50
For printer specifications, see page 3.
4-Terminal measurement probe for 3239 CLIP-TYPE LEAD 9287-10 (Approx. 85 cm between connectors, and 8 cm between probes) CLIP-TYPE LEAD 9452 (Approx. 80 cm between connectors, and 20 cm between probes) FOUR-TERMINAL LEAD 9453 (Approx. 80 cm between connectors, and 30 cm between probes) PIN-TYPE LEAD 9461 (Approx. 40 cm between connectors, and 25 cm between probes)
9287-10
9452
9453
9461
HIOKI (Shanghai) Sales & Trading Co., Ltd. :
1904 Shanghai Times Square Office, 93 Huai Hai Zhong Road Shanghai, P.R.China POSTCODE: 200021 TEL +86-21-6391-0090/0092 FAX +86-21-6391-0360 E-mail: info-sh@hioki.cn
DISTRIBUTED BY
HEAD OFFICE : 81 Koizumi, Ueda, Nagano, 386-1192, Japan TEL +81-268-28-0562 / FAX +81-268-28-0568 E-mail: os-com@hioki.co.jp HIOKI USA CORPORATION : 6 Corporate Drive, Cranbury, NJ 08512 USA TEL +1-609-409-9109 / FAX +1-609-409-9108 E-mail: hioki@hiokiusa.com
Beijing Office :
A-2602 Freetown, 58 Dong San Huan Nan Road Beijing, P.R.China POSTCODE: 100022 TEL +86-10-5867-4080/4081 FAX +86-10-5867-4090 E-mail: info-bj@hioki.cn Room 303, Profit Plaza, No.76, West Huangpu Road Guangzhou, P.R.China POSTCODE: 510623 TEL +86-20-38392673/2676 FAX +86-20-38392679 E-mail: info-gz@hioki.cn 3239E4-79M-03P Printed in Japan
Guangzhou Office :
All information correct as of Sep. 12, 2007. All specifications are subject to change without notice.


▲Up To Search▲   

 
Price & Availability of 3239

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .